Blue-White - Chemical metering and fluid measurement solutions; chemical metering pumps, peristaltic pumps, diaphragm pumps, multi-diaphragm pumps, flow meters, flowmeters, variable area flow meters, paddlewheel flow meters, ultrasonic flow meters, engineered skid systems, single skid system, duplex skid system, triplex skid system, water analyzers...
Oswald

Industrial

A-100NVP

Operation

Priming the Pump

Before priming the pump, or starting the pump, be sure that all suction and discharge lines are connected properly and that proper valves are open.  

Pressing the Run/Standby and Field buttons at the same time will start the pump at 100% speed and will run for 99 seconds.  The pump can be stopped at any time by pressing the Run/Standby button.  All 99 seconds may not be required to prime the pump.  Conversely, if additional priming is necessary, the process can be repeated.  

If the pump does not prime, check the suction line for clogs, obstructions, or leaks. Make sure all appropriate valves are open. 

Manual Operation – Mode 1

To run the pump in Manual Mode, ensure the “MODE 1” is displayed. Or toggle Mode button until “MODE 1” is displayed. Press the Run/Standby key to start the pump running. Adjust pump speed by programming (see Set-Up and Controls section). To stop the pump, press the Run/Standby key.  

4-20ma Input Speed Control – Mode 2

To run the pump in 4-20mA Mode, press the Mode button until “MODE 2” is displayed. Press the Run/Standby button to start the pump.  Pump speed will be set by remote 4-20mA input signal.  To toggle/view either % motor speed or 4-20 mA signal, press the Digit and Mode keys at the same time. To stop the pump, press the Start/Stop key.  

0-10 VDC Input Speed Control – Mode 3                                        

To run the pump in O-10 VDC Mode, press the Mode button until “MODE 3” is displayed. Press the Run/Standby button to start the pump.  Pump speed will be set by remote 0-10 VDC input signal.  To toggle/view either % motor speed or 0-10 VDC signal, press the Digit and Mode keys at the same time. To stop the pump, press the Start/Stop key.  

Frequency (Hz) Input Speed Control – Mode 4

To run the pump in Frequency Mode, press the Mode button until “MODE 4” is displayed. Press the Run/Standby button to start the pump.  Pump speed will be set by remote Hz input signal.  To toggle/view either % motor speed or Hz signal, press the Digit and Mode keys at the same time. To stop the pump, press the Start/Stop key.  

Calibration

To ensure the most accurate pumping, a pump calibration is recommended prior to operating the pump.  The A-100NVP pump flow rate will vary depending on the suction line conditions. Calibrate the pump with actual suction line conditions.     

To calibrate the pump: 

  1. Connection tubing and open valves on the suction side of the pump to represent actual suction lines conditions. 
  2. Set up discharge and open discharge valves to dispense into calibration column/cylinder. 
  3. Prime to the pump so that the calibration column begins to fill. Stop the pump and note the solution level in the calibration column.  
  4. Run the pump at a set speed for 1 minute.   
  5. Record the flow difference in the calibration cylinder.  
  6. Calculate the flow rate by dividing the flow by the time. For example, if the pump runs for 1 minute, and the flow difference is 30 ml, then divide 30ml by 1 minute. The flowrate is 30 ml/min.   
  7. For best accuracy, run more than one calibration to check results, or run another calibration at a different pump speed.

Calibration results should be similar to the flow rating for the pump tube, but can vary depending on chemical composition, suction lift, viscosity, temperature, and other conditions.   

Tube Failure Detection (TFD+)

image 17

The A-100NVP pump is equipped with a corrosion-resistant TFD+ leak-detection located inside the pump head.  If a tube fails or leaks for any reason, the polymer or chemical will trigger the sensor, stop the pump, and activate an alarm.  

If TFD+ alarm occurs, the pump will stop, close an alarm output, and the screen will display “Alarm”. The TFD+ can detect the presence of non-conductive chemicals, such as oil-based polymers, water-based polymers, or other oils.  

If the system has detected a leak, the pump tube must be replaced and the pump head and roller assembly must be thoroughly cleaned. Remove the front cover to replace the tube and roller. Failure to clean the pump head properly will void the warranty.  

See video on proper tube replacement.  https://www.youtube.com/watch?v=NqhrR2Vs1OI&t=69s 

Confirm Detection

To determine if your chemical will be detected by the system, remove the pump tube and roller assembly. Place a small amount of the chemical in the bottom of the pump head – just enough to cover the sensor.  
Turn on the pump – If the TFD+ system detects the chemical, the pump will stop after a five-second confirmation period and the ALARM icon will light on the display. If the TFD+ system does not detect the chemical, the pump will continue to run after the confirmation period.  
Carefully clean the chemical out of the pump head, being sure to remove all traces of chemical from the sensor probes. Press the RUN/STAND-BY button to clear the alarm condition and restart the pump.  
Contact Closure Alarm Output – A contact closure output (relay) is provided with the TFD+ system. The relay can be configured for normally open (factory default) or normally closed operation by properly positioning the connector plug on the circuit board. 

FAQs

Frequently Asked Questions

A-100NVP Operation

  • How do I reset the 500 hour service timer?

    Refer to page 9 and page 18 in the manual. Press and hold run/standby button & digit buttons at the same time.

  • Will pump display incoming milliamp values?

    Yes, when in mode 2, press the digit & mode button to view the incoming mA values and corresponding motor speeds.

Please click here to see help center disclaimers.

  • Ask me a question!
Processing... ...

Stay in the know

Request a quote

Please fill out the form to request a quote.
A sales rep will reach out to you.

To select or deselect multiple products–desktop users hold [ctrl] on Windows or [cmd] on Mac and click products.

Select Products

Pumps
Flow Meters
Skid Systems
Analyzers

Please review your selections before continuing.

Product Details
Select your industry:
How did you hear about us?

We'd like to hear from you...

ImageCatalog NumberDescriptionPriceBuy