Key Features:
- Turn-Key Monitoring Solution for Clean Water Applications
- Highly Accurate, Real-Time Measurement, Display, and Data-Logging
- Ultra-Low Turbidity, Free or Total Chlorine, pH, and Temperature
- Touch Screen Display and Data-Logging Terminal
- FCL700 Smart Sensor
- UT700 Ultra-Low Turbidity Sensor with Unique Flat Surface Distal
End in Quartz Glass Plate
Technical Overview:
pH Method
Electrochemical Method
Sensor
Flat Electrode
Sampling Chamber
Dual Sensor Flow Reservoir
Warranty
13 Months
Free Chlorine Method
Bare Gold - Electrochemical Method, Chlorine Range: 0.1–5 PPM
Turbidity Dual Range
.001 - 10 / 10 - 40.00 NTU
Turbidity Accuracy
+/- 0.005 NTU or 2% <10NTU
Turbidity Repeatability
+/- 0.001 NTU or 0.5% <10NTU
Literature
APH2O Literature
Instruction Manuals
APH2O Instruction Manual – 80000-186Technical Data Sheets
APH2O Technical Data Sheet – 85000-186
3D Visualizer
Features
Overview
- Ultra-low resolution UT700 turbidity sensor offers a detection light source using warm white LED in 90-degree surface scatter format in accordance with USEPA 180.1 standards.
- FCL700 three-parameter composite sensor for residual chlorine, pH, and temperature in compliance with USEPA 334.0 guidelines. Advanced sensor PCB offers built-in temperature and pH parameter compensation (up to pH 9.0) algorithms.
- Unique Bare-Gold electrode technology for residual chlorine measurement eliminates membranes and electrode solution replenishment commonly associated with conventional sensors. Flat bubble pH electrode design for reduced fouling potential.
- Dual-Sensor flow reservoir provides sample calming for dissipation of air bubbles and settling of suspended solids, foam or other impurities commonly observed in drinking water influent. Allows the highest level of turbidity resolution on the market, and greatly extends the maintenance cycle of the sensor while providing a large buffer capacity to mitigate pressure fluctuations.
Applications
- Drinking Water
- Domestic Water
- Secondary Water Supply
- Sanitary Water
- Decorative Water / Swimming Water
- Food & Beverage Process Water
What’s Included
- One APH2O Drinking Water Panel Assembled and Complete with 110VAC Power Supply Cord
- TS10 Touch Screen display/ Data Logger with Sensor pre-wire in RS-485 (RTU).
- WR20 Dual Sensor Flow Reservoir
- Ultra Low Turbidity Sensor
- Chlorine Sensor
- Inlet Flow meter
- Isolation Valve
- Inlet Tubing
- Outlet Tubing
- Rear FRP channel strut (3) and hardware
- Electrical Schematics
Specifications
Item
- APH2O
Turbidity Wavelength
- Warm White
Light Source
- LED
Turbidity Dual Range
- 0.001 – 10 / 10 – 40.00 NTU
Turbidity Accuracy
- +/- 0.005 NTU or 2% <10NTU
Turbidity Repeatability
- +/- 0.001 NTU or 0.5% <10NTU
Free Chlorine Method
- Bare Gold – Electrochemical Method
Chlorine Range
- 0.01 – 5.00 ppm as FCl2 (Free) or TCl2 (Total)
pH Method
- Electrochemical Method
pH Range
- 0.01 – 14.00 pH
Measurement Accuracy
- 0.001 NTU or ± 1% Full Scale
- +/- 0.01 ppm or 1% Full Scale w/pH compensation to 9.0
- +/- 0.01 pH
Minimum Resolution
- 0.001 NTU
- 0.01 ppm
- 0.01 pH
Response time
- 4s after immersion – Turbidity
- T95≤60s – Free Chlorine
- T95≤5s – pH
Compliance
- EPA-180.1/334.0
Measurement Interval
- Continuous Measurement
Display
- 7-inch LCD Color Industrial Capacitive Touch Screen
Storage Capacity
- Built-In 4GB of Ram for Storing up to 1-Million Data/Event Records
Power Requirement
- 96-260VAC / 50-60 Hz; 10A Fuse; 200 W
Output
- 3 x 4-20 mA / RS-485 Modbus – RTU / Modbus TCP
Input
- 3 x 4-20 mA / RS-485 Modbus – RTU
USB
- 1 x USB host, for data downloading and screen upgrade
Internet
- RJ-45 socket, Modbus-TCP
Panel Operational Temperature
- 40 – 113oF (4-45 oC)
Storage Temperature
- Instrument: -4 – 131oF (-20 – 55oC) / Sensors 32 – 122oF (0 – 50oC)
Sample Water Temperature
- 40 – 104oF (4-40oC)
Sample Water Pressure
- 7.25 – 30 psi (0.05 – 0.2MPa)
WR20 Sample Water Flow Rate
- 9.5 – 15.5 GPH (600-1,000mL /Minute)
WR20 Line Size
- 1/4” Inlet / 3/8” Overflow / 3/4” Outlet / 1/2” Drain
Rating
- IP-65 Panel-Display / IP-67 Sensors
Regulation
- CE / RoHS
Relative Humidity
- 20% – 90% (No Condensation)
Altitude
- <6,561 feet (<2,000 Meter)
Dimensions (HxWxD)
- Panel 36 H x 20 W x 9.42 D inches
Approximate Product Weight
- 45 lbs
Videos
Accessories
Related Articles & Case Studies
Downloads
Title | Size | Update Date | Download |
---|---|---|---|
APH2O Instruction Manual – 80000-186 |
2.32 MB | September 18, 2024 | Download |
APH2O Quick Start Guide – 80000-189 |
392.68 KB | April 16, 2024 | Download |
APH2O Technical Data Sheet – 85000-186 |
708.84 KB | September 12, 2024 | Download |
Municipal Catalog 2024 – 80000-708 |
13.09 MB | July 16, 2024 | Download |
Municipal Products Brochure |
2.98 MB | February 14, 2024 | Download |